31JD151 |
1JD152 |
2JD153 |
3JD154 |
4JD155SD/Aylin/others to measure trees and collect cores (start) |
5JD156 |
6JD157
|
7JD158at ~1600 MDT, the 2.5 m LI-7500A stopped working.
|
8JD159SD/Aylin/others to measure trees and collect cores (end)
|
9JD160
|
10JD161
|
11JD162In lab (HR) prep for site visit
|
12JD163Site Visit (HR, SpB): Details , SpB Photos (35) , HR Photos (7) AP200 N2 cylinder: 600 psi from 10:00-10:30 MST, removed 2.5-m REBS Rnet sensor and LI-7500A. Updated UCB programming to remove REBs and snow depth measurements at ~15:00 MST removed fallen tree from NE tower guy wire and re-measured tension (~900 lb pre-tree removal, ~425 lb post-tree removal) Downloaded all data except Bowling logger. Turned off NOAH logger. Downloaded data and replaced batteries in zentra logger. From ~13:30-15:00 MST, swapped out LI-7200s. SN 72H-0539 is now on the tower and SN 72H-1033 is back in Boulder to be sent back to LBL. from 15:35-15:45 MST, adjusted AP200 N2 flow rate at around 15:20 MST, started ops120 and ucb_44.csi; removed Rnet.2m, removed snow depth (D.judd, T.judd), added back SN-500 Rnet (Rsw.in.sn500 Rsw.out.sn500 Rlw.in.sn500 Rlw.out.sn500) HR Jeep Partly cloudy and windy
|
13JD164at around 14:30 MST, HR rebooted the 21.5m LI-7200 (added back variable "Diagnostic Value 2")
|
14JD165
|
15JD166
|
16JD167
|
17JD168at around 9:30 MST, HR rebooted the 21.5m LI-7200 (now saving 30-min data files rather than hourly files)
|
18JD169
|
19JD170Site Visit (HR): Details , HR Photos (9) from 12:50-13:30 MST, calibrated LI-7500A (sn: L5H-2499) and reinstalled on 2.5m tower Downloaded data from Bowling logger HR Jeep Sunny and calm
|
20JD171
|
21JD172Summer Solstice
|
22JD173at around 13:45 MST, the 21.5m LI-7500RS/A (op2) rebooted (likely power outage) SpB Off
|
23JD174
|
24JD175
|
25JD176
|
26JD177
|
27JD178
|
28JD179
|
29JD180at around 10 MST, the CNR4 logger stops sending data to the RAD/CNR logger (suspect a power/battery issue) SpB Returns
|
30JD181
|
1JD182
|
2JD183
|
3JD184CU Holiday (Day before Independence Day)
|
4JD185At 15:25 MST, turned off LI-7200 fan since flow rate was ~5 slpm Independence Day |