27JD300 |
28JD301at around 10 MST, 2.5m LI-7500 went off-scale? (check this!) |
29JD302 |
30JD303 |
31JD304Halloween Windy (boo!) |
1JD305 |
2JD306Site Visit (SpB): Details , Photos (28) Span cylinder at 1210 psi. N2 at 1950 psi swapped LI-7200s (sn 72H-0479 replaced sn 72H-0192), LI-7550 box, and Fan Module (LI-7200-101) reduced LI-7200 inlet heater from 5W to 4.6W couldn't connect to 2.5m LI-7500 (serial port problem?) CU Jeep Sunny and breezy. More snow on ground. |
3JD307 |
4JD308at 10:40 MST, changed LI-7200 flow rate from 13 SLPM to 11 SLPM, also added site setup info |
5JD309 |
6JD310at 12:00 MST, reduced the LI-7200 flow rate from 11 SLPM to 10.8 SLPM |
7JD311Site Visit (SpB, EG, M): Details , Photos (27) Span cylinder at 1190 psi. N2 at 1800 psi started ops68 (renamed UC CSATs to correct hts, added 2D ATI), new UC logger program (uc_05.csi). swapped LI-7200 heated inlet, reduced power from 4.6W to 4W 2.5m LI-7500 (AGC problem?) from approx 16:06-16:15 MST, Changed profiler (li-6251) dessicant SpB Jeep Sunny, fairly warm. Wave clouds. |
8JD312Between 10-12:30 MST, many restarts to LI-7200 data system (looks like multiple power outages)? Windy. |
9JD313 |
10JD314 |
11JD315 |
12JD316Site Visit (SpB, JH, SM): Details , Photos (11) Span cylinder at 1150 psi. N2 at 1700 psi added analog out from CSAT3 into LI-7550, moved LI-7200 inlet closer to CSAT sampling volume (calculate new separation), returned LI-7200 cables to Stefan confirmed 2.5m LI-7500 problem (appears to be related to the connector) Sunny, warm, calm. |
13JD317 |
14JD318 |
15JD319Site Visit (ES): Details at 11:05 MST, updated LI-7200 site setup info with correct North/East separation distances In afternoon: ES added a battery to the CPEC200 system |
16JD320 |
17JD321 |
18JD322at 10:10 MST, changed LI-7200 sample rate from 20 samp/s to 10 samp/s, also changed bandwidth from 10 Hz to 5 Hz |
19JD323 |
20JD324 |
21JD325Site Visit (TB, EB): Details started mux replacement at N Canopy tower |
22JD326 |
23JD327Site Visit (TB): Details finished mux replacement at N Canopy tower |
24JD328at 11:50 MST, changed LI-7200 sample rate from 10 s/s to 20 s/s (bandwidth from 5 Hz to 10 Hz); also changed range for DAC co2 dry to 380-430 umol/mol |
25JD329 |
26JD330 |
27JD331 |
28JD332TURKEY DAY |
29JD333Site Visit (SpB): Details , Photos (23) Span cylinder at 1000 psi. N2 at 1300 psi replaced 2.5m LI-7500 (sn 75H-0082) with PB spare (sn 75H-0062) covered small gap in LI-7200 inlet tube with pipe insulation started using a new FAST logger program (fast_21.csi, note fast_20.csi was skipped) to try and speed up fast logger. CU Prius Sunny and breezy. |
30JD334 |